Description
Permission RCM markc UL us – Listed (OL)
CE certificate (see declaration of conformity) acc. to EU-EMV Directiveacc. to EU-RoHS Directive
KC-mark KC-EMV
Material information RoHs compliant
Measured parameter flowtemperature
Flow direction P1 -> P2Uni-directional
Measurement process temperature: PT1000Flow: Vortex
Flow measurement range starting value [l/min] 1,8
Flow measurement range end value [l/min] 32
Operating pressure [bar] 0 to 12
Information about operating pressure max. 12 bar at 40??Cmax. 6 bar at 100??C
Operating medium Fluid mediaNeutral liquidsWater
Information about operating and control medium Media compatibility to the media contacting materials must be guaranteed
Medium temperature [??C] 0 to 90
ambient temperature [??C] 0 to 50
Nominal temperature [??C] 23
Accuracy flow value ?I2 %FS for flow <= 50 %FS?I 3 % o.m.v. for flow >= 50 %FS
Accuracy temperature in ?I ??C [??C] 2
Repeatability flow value < ?I0.5 %FS for flow <= 50 %FS< ?I1 % o.m.v. for flow >= 50 %FS
Temperature coefficient clamp in ?I %FS/K type ?I0,05%FS/K
Control output 2 x PNP or 2 x NPN switchable
Switch function Window comparatorThreshold comparatorFreely programmable
Switching element function Opener/closer switchable
Max. output current [mA] 100
Analogue output 0 – 10 V4 – 20 mA1 – 5 V
Flow characteristic curve starting value [l/min] 0
Flow characteristic curve end value [l/min] 32
Max. load resistance current output [Ohm] 500
Min. load resistance voltage output [kOhm] 15
Short circuit resistance yes
Overload resistant Available
Report IO-Link
IO-Link, report Device V 1.1
IO-Link, profile Smart sensor profile
IO-Link, function classes Teach channelDiagnosisIdentificationBinary data channel (BDC)Process data variables (PDV)
IO-Link, communication mode COM2 (38.4 kBaud)
IO-Link, SIO mode support yes
IO-Link, port type A
IO-Link, process data width OUT 0 Bytes
IO-Link, process data width IN 5 Bytes
IO-Link, process data contents IN 1 bit BDC (volume monitoring)14 bit PDV (flow measurement)1 bit BDC (temperature monitoring)14 bit PDV (temperature measurement)2 bit BDC (flow measurement)
IO-Link, service data unit IN 32 bit Volumenmesswert
IO-Link, minimum cycle time 5 ms
IO-Link, data memory needed [Byte] 500
Operating voltage range DC [V] 18 to 30
Reverse polarity protection for all electrical connections
Electrical connection M12x15 pinPlug straightA-coded
Max. line length 30 m20 m For OK link operation
installation position arbitrary
Fluid connection Hose screw connection 13 mm
Product weight [g] 300
Housing material PA reinforced
From medium contacting materials EPDM (perox.)ETFEstainless steelPA6T/6I reinforced
Presentable unit(s) US galUS gal/mincftcft/minll/hl/minm3??C??F
Protection category IP65
Corrosion resistant class KBK 3 – severe corrosion stress
Temperature measurement range stating value [??C] 0
Temperature measurement range end value [??C] 90
Temperature characteristic curve starting value [??C] 0
Temperature characteristic curve end value [??C] 100
Reviews
There are no reviews yet.